DZDR-P100 is a newly launched flat plate thermal conductivity tester by Nanjing Dazhan Instrument, which adopts the single specimen protection hot plate method (GHP). The instrument is equipped with a 300 × 300mm standard specimen table, an automatic pressure thickness measurement system, and a high-precision temperature control module. It can achieve thermal conductivity of 0.002-2.0W/(m · K), wide range measurement, and repeatability of ± 1%. It is an ideal equipment for building energy conservation, insulation material research and quality control.
The Guarded Hot Plate Method (GHP) is a benchmark method for determining the thermal conductivity of insulation materials. DZDR-P100 adopts a single specimen measurement structure, placing the specimen between the hot plate and the cold plate, and using a protective plate surrounding the hot plate to eliminate lateral heat loss. After the system reaches thermal steady state, the thermal conductivity λ is calculated by measuring the heating power P, temperature difference Δ T, specimen thickness d, and area A according to Fourier's law
(Formula: λ=P × d/(A × Δ T)), and the thermal resistance R=d/λ can be further derived. The uniform temperature design of the protective plate effectively reduces edge heat loss, ensuring the stability and reliability of test data.
The DZDR-P100 flat plate thermal conductivity tester is widely used for testing the thermal conductivity of materials in fields such as building energy conservation, insulation materials, aerospace, automotive, and home appliances.
Features
Wide range measurement: support the range of 0.002~2.0W/(m · K), covering a variety of materials from aerogel felt to bakelite board.
High precision temperature control: The temperature control accuracy reaches ± 0.02 ℃, with a repeatability deviation of<1%
Standardized configuration: equipped with a 300 × 300mm standard specimen table, in compliance with national standards and engineering acceptance specifications.
Fully automated operation: integrating the full process automation of pressurization, thickness measurement, testing, and report generation, simplifying the operation steps.
Strong compliance: Complies with international and domestic standards such as GB/T10294, ISO8302, ASTMC177, etc.
Model Expansion: Optional High Temperature P100A (Hot Plate Temperature 150-200 ℃) to meet a wider range of testing needs.
Parameters
| Technical Specifications | DZDR-P100 |
| Test method | Single specimen protection hot plate method (GHP) |
| Thermal conductivity range | 0.001 ~ 2.0 W/(m·K) |
| Thermal resistance range | 0.01 ~ 4 m²·K/W |
| Repeatability/accuracy | Repeatability ± 1%; Accuracy routine ± 3%, standard substance calibration ± 2% |
| temperature range | Hot plate RT+10~100 ℃, cold plate -5~90 ℃; Optional high-temperature hot plate 150-200 ℃ |
| temperature control accuracy | ±0.02℃; Temperature resolution 0.01 ℃ |
| Sample Specifications | 300×300×(5~50) mm; The flatness of hard materials is better than 0.025% |
| Pressure/thickness measurement | Pressure can be continuously adjusted from 0 to 4 kPa, with a resolution of 0.1 kPa; Automatic thickness measurement from 0 to 50mm, with an accuracy of 0.005mm |
| Communication output | USB+backup RS232, automatic test/report output; Optional wireless connection |
| environmental conditions | Environmental temperature 0~40 ℃, relative humidity ≤ 65% RH |
Related Products
Inquiry
Nanjing Dazhan Testing Instrument Co., Ltd., located in Jiulong Lake, Jiangning District, Nanjing, is a high-tech enterprise integrating scientific research, production, and sales.